Shannon Dulz
AST 311
September 30, 2015
Detector Lab Report
Abstract
In this lab, we obtained flat, dark
and bias images on the CCD cameras in order to characterize background levels
of these CCDs for later processing of images. We calculated the dark current
rate and average bias levels.
Introduction
As part of every astronomical image
obtained with a CCD, there are background counts of electrons resulting from
the detector itself. To account for these background levels, three types of
calibration images are taken: bias, darks and flats. Bias images account for
noise introduced from the readout of the detector. Biases are taken as
zero-second exposures with a closed shutter. Dark images account for the
accumulation of noise due to the electronics of the detector and it tends to
grow linearly over the integration time. Dark images are taken at the same exposure
length as flat images (or scaled to the same length) and with the shutter
closed. Flat images account for pixel-to pixel variations in the detection of
light. Flat images are taken of an illuminated screen (or the sky at twilight
for sky-flats) with the shutter open at an exposure length results in the image
being evenly illuminated but not yet saturated. During processing, the bias
frames are averaged into a master bias and subtracted from all the other images
including darks and flats. Then, the darks are averaged into a master dark,
which is subtracted from the science images and the flats. Lastly the flats are
averaged into a master flat, which is divided from the science image
pixel-by-pixel.
Procedures
For this lab, we tested the CCD cameras
without a telescope. The cameras were connected to our laptops via USB cables.
Then the program Maxim DL was used to collect images. To connect the camera in
the program, under the settings tab on the program, we clicked set-up camera
and insured the correct camera model is shown. Also insure, swap chips is off,
guide chips is internal, and ext trigger is off. Under the options tab, turn
off rotation orientation and auto dark subframe extraction. Finally, click
connect.
To begin taking images, under the
expose tab, adjust the exposure time to required time. Images must be saved
after each exposure unless Auto Save is used. We began by taking 10 bias images
of exposure time zero seconds with the cap on the detector to block light. One
of the bias images is below.
Next,
we obtained flat fields with the camera pointed at an illuminated piece of
paper. We took several test images at various exposure times to insure counts
would be higher than bias levels but not saturate the detector. We choose an
integration time of 0.01 seconds and took 10 flat images, an example of which
is shown below.
While
this image, appears identical to the bias image, the counts are in fact much
higher and a slight grating in counts from top to bottom is noticeable. Using
this same integration time of 0.001 seconds we obtained 10 dark images with the
cap on the camera, an example of which is shown below.
Lastly,
in order to calculate dark current levels we obtained 1 image each at 1, 5, 10
and 20 second exposure times with the cap on the camera.
Results and Discussion
For
this lab, we obtained multiple flats, darks, and bias (shown in the table
below).
|
Image
Type
|
Integration
time (seconds)
|
Number
of Images
|
|
Bias
|
0
|
10
|
|
Flat
|
0.01
|
10
|
|
Darks
|
0.01
|
10
|
|
Darks
|
1,
5, 10, 20
|
1
each
|
Using IDL, the biases were averaged
together by summing the counts in each pixel of each image and dividing that
number by the total number of images. This produced a master bias frame the
median of which was 2006, giving us the average bias count. The darks of
exposure time 0.01 were averaged in the same way leading to an average dark
count for those images of 1789. Lastly the flat fields were averaged into a
master flat, the median of which was 4299. The average counts of the biases
should not normally be higher than that of the darks. This variable bias level
could be due to temperature and we will recalculate the bias level from the
dark current rate discussed below. While bad pixels were evident in individual
images by vastly different count levels in single pixels, these few dozen
pixels averaged out when the images were summed together. The flat fields we
obtained were very evenly illuminated with only a slight grating of counts from
top to bottom of the image. To obtain the dark current rate, we obtained darks
at 1, 5, 10 and 20 seconds. We took the median of these images individually to
obtain the table below.
|
Exposure
time (seconds)
|
Average
counts
|
|
1
|
1026
|
|
5
|
1031
|
|
10
|
1037
|
|
20
|
1036
|
We
expect to see the dark counts increase linearly over time, the slope of which
would give the dark current. While the linear fit is far from perfect due to
the few number of images taken, we obtained a dark current of 0.5 counts/sec.
The y-axis intersection of this linear fit is the bias level for these sets of
images. For this we obtain 1028 counts as the bias level. This calculation is a much more reasonable
number than that obtained from the bias images themselves of 2006. This
discrepancy could be due to temperature, especially since the bias images were
the first ones taken as the CCD was still warming up. The graph of this linear
fit is below.
Conclusions
In this lab, we took flat, dark and
bias images to begin to characterize the CCD detectors. We found an average
bias level over 10 images of 2006 counts. When recalculated using the linear
fit intersection of the dark current, we obtained a much more reasonable 1028
count bias level. We obtained a very low dark current rate of 0.5 counts/sec.
Lastly, we took very even flat field images with average counts of 4299. The
low background levels of these detections especially with regard to the low
dark current will allow for high signal to noise ratios when they are used for
later labs.




No comments:
Post a Comment